Giving Shape to Ideas

search form

Specifications

Measuring Instruments

CS-200

Luminance & Color Meters

The CS-200 ChromaMeter measures luminance and chromaticity of light emitting products as diverse as large plasma displays, compact LCDs and LEDs, outdoor screens, high pressure lamps, and instrument panels.

Specifications

Principal Specifications CS-200

ModelCS-200
Measurement range0.01 - 200,000 cd/m2 (Measuring angle 1°)
0.01 - 5,000,000 cd/m2 (Measuring angle 0.2°)
0.01 - 20,000,000 cd/m2 (Measuring angle 0.1°)
Accuracy
(Measuring angle 1°) #1
(Temperature: 23°C±2°C, Relative humidity: 65% max.)
150 cd/m2 ( for Illuminant A ) : Lv ±2 % ±1digit ; xy ±0.002
0.01-0.5 cd/m2 ( for Illuminant A ) : Lv ±0.02 cd/m2 ±1digit ;
0.5-1 cd/m2 ( for Illuminant A ) : Lv ±0.02 cd/m2 ±1digit ; xy ±0.007
1-10 cd/m2 ( for Illuminant A ) : Lv ±2 % ±1digit ; xy ±0.004
10-200,000 cd/m2 ( for Illuminant A) : Lv ±2 % ±1digit ; xy ±0.003
5000 cd/m2 ( for Illuminant A ) : color filter (R, G, B) ; xy ±0.006
Repeatability
(Measuring angle 1°) #2
0.01-1 cd/m2 ( for Illuminant A ) : Lv ±0.01 cd/m2 ±1digit ;(2σ/AUTO)
1-2 cd/m2 ( for Illuminant A ) : Lv ±0.5 % ±1digit ; xy 0.002 (2σ/AUTO)
2-4 cd/m2 ( for Illuminant A ) : Lv ±0.5 % ±1digit ; xy 0.001 (2σ/AUTO)
4-8 cd/m2 ( for Illuminant A ) : Lv ±0.5 % ±1digit ; xy 0.0005 (2σ/AUTO)
8-200,000 cd/m2 ( for Illuminant A ) : Lv ±0.1% ±1digit ; xy 0.0004 (2σ/AUTO)
Measurement timeAutomatic setting between 1s and 60s (AUTO); Automatic setting to 1s or 3s (LTD. AUTO)
0.5 sec / meas. (Super-FAST) 1 sec / meas.(FAST)
3 sec / meas. (SLOW) 12 sec / meas. (Super-SLOW)
Measurement methodSpectral method, Grating + linear photo diode array
Measuring angle1°, 0.2°, 0.1°(switchable)
Minimum measuring area0.5 mm
0.1 mm (close up lens)
Minimum measuring distance296 mm (Distance from front edge of metal lens barrel)
Observer2/10 degrees
Color spaceLv x y, Lv u' v', Lv TΔuv, XYZ, dominant wavelength
Measurement synchronization setting rangeVertical synchronization frequency : 40.00 to 200.00Hz
InterfaceUSB 1.1
Power sourceAC adapter or 4 AA-Size Batteries
Battery lifeApprox. 3 hours (continuous measurement / Fast mode / AA-size alkaline cells)
Size95 mm (W) × 127 mm (H) × 330 mm (L)
Weight1.8 kg (without battery)
Operating temperature/humidity range0°C to 40°C, relative humidity 85% or less (at 35°C) with no condensation
Storage temperature /humidity range0°Cto 45°C,relative humidity 85% or less (at 35°C)with no condensation
#1
23°C±2°C Lv = 0.01-10 cd/m2, SLOW, average of 30 measurements
Lv = 10 cd/m2 and higher, SLOW, average of 10 measurements
#2
At 0.2°measuring angle, the amount of received light is approx. 1/25 of that for 1°.
Therefore, the repeatability becomes the same as that for 1°with 25 times lower luminance.
At 0.1°measuring angle, the amount of received light is approx. 1/100 of that for 1°.
Therefore, the repeatability becomes the same as that for 1°with 100 times lower luminance.  
*
Specifications are subject to change without prior notice.

Sus

© 2012 Konica Minolta